Organ at risk
|
Volume (cm3) | Volume maximum (Gy) | Point dose maximum (Gy) |
| Spinal cord | 0.1 | 25 | 30 |
| <0.35 | 23 | ||
| <1.2 | 14.5 | ||
| Liver−GTV | 700 | 15 | |
| Mean < 14 Gy | |||
| Stomach | 0.1 | 27.5 | 32 |
| <10 | 18 | ||
| Duodenum | 0.1 | 30 | 32 |
| <5 | 18 | ||
| <10 | 12.5 | ||
| Esophagus | <5 | 19.5 | 35 |
| Small bowel | <5 | 19.5 | 35 |
| Large bowel | <20 | 25 | 38 |
| Lung total | 1500 | 12.5 | |
| Lung total | 1000 | 13.5 | |
| Kidney total | 200 | 17.5 | |
| Heart/pericardium | <15 | 32 | 38 |
| Chest wall/rib | <30 | 30 | 43 |
| Skin | <10 | 36.5 | 39.5 |
W. Tristram Arscott, Reid F. Thompson, Lingshu Yin, Brendan Burgdorf, Maura Kirk, Edgar Ben-Josef,
Stereotactic body proton therapy for liver tumors: Dosimetric advantages and their radiobiological and clinical implications,
Physics and Imaging in Radiation Oncology,
Volume 8,
2018,
Pages 17-22,
ISSN 2405-6316,
https://doi.org/10.1016/j.phro.2018.11.004.
(http://www.sciencedirect.com/science/article/pii/S2405631618300484)
Example case shown planned with this RapidPlan model
|
Supporting Documents |
Download RapidPlan model |
DICOM patient export |
Any reference to a "plan study" are simply what the organizers call each case and may not be a "study" in the FDA sense as they may not have been published in a peer reviewed journal.
Varian does not provide medical advice and these are illustrative examples only.
Leading plans by expert planner. Your results may vary.
FOR EDUCATIONAL AND SCIENTIFIC EXCHANGE ONLY – NOT FOR SALES OR PROMOTIONAL USE.