VMAT

Rx: 16Gy/1fx

 

Beams
DVH
Simple DVH Metrics

Name (ID): AVM, AVM (033020172)
Plan or PlanSum ID: Edge

Structure ID Structure Code Patient Structure DVH Objective Evaluator Variation Priority Met Achieved
3.2GyBody   BODY V3.2Gy[cc]         50.39 cc
9.6GyBody   BODY V9.6Gy[cc]         11.23 cc
16GyBody   BODY V16Gy[cc]         4.95 cc
Body   BODY Mean[Gy]         0.429 Gy
PTV1   GTVp V100%[%] >=99     Goal 99.20 %
PTV1   GTVp Max[%]         133.8 %
PTV1   GTVp Max[Gy]         21.407 Gy
PTV1   GTVp Min[Gy]         14.719 Gy
PTV1   GTVp Volume[cc]         4.52 cc
Technical Plan Comments

This plan was created from an MR dataset as that is all that was available.  Dose was calculated by overriding the density of the dataset to that of water.  One full VMAT arc was used and 5 half arcs were added, each with the couch rotated an additional 30 degrees.  The goal of this plan was to keep the 20% and 60% isodose levels as compact as possible.  The plan was optimized with a single lower dose objective on the target, the manual NTO and single ring//shell structure penalizing dose beyond 20% (3.2Gy). The extended convergence calculation option was enabled for the Photon Optimizer, available in Eclipse 15.5+.

Links

3rd party software plan report

PDF

                              

DICOM patient export     

ZIP        

Any reference to a "plan study" are simply what the organizers call each case and may not be a "study" in the FDA sense as they may not have been published in a peer reviewed journal.
Varian does not provide medical advice and these are illustrative examples only.
Leading plans by expert planner. Your results may vary.

FOR EDUCATIONAL AND SCIENTIFIC EXCHANGE ONLY – NOT FOR SALES OR PROMOTIONAL USE.