Name (ID): Liver, SBRT (LiverSBRT_med___affairs)
Plan or PlanSum ID: Edge
Structure ID | Structure Code | Patient Structure | DVH Objective | Evaluator | Variation | Priority | Met | Achieved |
PTV | P_R_liver_50 | V50.0Gy[%] | >=95 | 90 | Goal | 99.51 % | ||
LIVER-GTV | O_normal_liver | Mean[Gy] | <=13 | 13.2 | Goal | 12.040 Gy | ||
ESOPHAGUS | O_Esgs | D0.05cc[Gy] | <=32 | 34 | Goal | 17.492 Gy | ||
STOMACH | O_Stmc | D0.05cc[Gy] | <=30 | 32 | Goal | 8.868 Gy | ||
DUODENUM | O_bowel_bag | D0.05cc[Gy] | <=30 | 32 | Goal | 5.280 Gy | ||
SMALL_BOWEL | O_bowel_bag | D0.05cc[Gy] | <=30 | 32 | Goal | 5.280 Gy | ||
LARGE_BOWEL | O_bowel_bag | D0.05cc[Gy] | <=32 | 34 | Goal | 5.280 Gy | ||
SPINAL_CORD_05 | SpinalCord_05 | D0.05cc[Gy] | <=25 | 28 | Goal | 19.067 Gy | ||
KINDEYS | Kidneys | Mean[Gy] | <=10 | 12 | Goal | 1.606 Gy | ||
LIVER-GTV | O_normal_liver | D70.0%[Gy] | <=70 | Goal | 1.829 Gy | |||
LIVER-GTV | O_normal_liver | V10.0Gy[cc] | <=700 | Goal | 448.18 cc |
Two arc (2) VMAT technique was chosen and 2 almost complete noncoplanar arcs were used with the couch rotated 13 degrees eityher direction from zero. Unique collimator rotations per field were selected (315/45). By far, the most challenging aspect of this plan was achieving the mean dose to uninvolved liver constraint. By placing large constraints on healthy liver in the optimizer that constraint was able to be met. Do to the non coplanar arrangement lateral dose spill was minimized. The photon optimizer used the extended convergence calculation option and the MU objective was enabled (strength 70) with minimum MU 1000 maxiumum MU 3000. All constraints were met as requested by the protocol with no variations cited.
3rd party software plan report |
|
DICOM patient export |
Any reference to a "plan study" are simply what the organizers call each case and may not be a "study" in the FDA sense as they may not have been published in a peer reviewed journal.
Varian does not provide medical advice and these are illustrative examples only.