3:43 IMRT treatment field delivery time
Name (ID): ROQS-22, HN(PinPC2013) (ROQS-22)
Plan or PlanSum ID: sf9smKVm5IDS1
Structure ID | Structure Code | Patient Structure | DVH Objective | Evaluator | Variation | Priority | Met | Achieved |
PTV7000org | PTV7000org | V70.0Gy[%] | >=95 | 90 | Goal | 95.79 % | ||
PTV7000org | PTV7000org | V66.5Gy[%] | >=99 | 97 | Goal | 100.00 % | ||
PTV7000org | PTV7000org | D0.03cc[Gy] | <=72.8 | 77 | Variation | 73.198 Gy | ||
PTV_6300_EVAL | PTV_6300_EVAL | V63.0Gy[%] | >=95 | 90 | Goal | 96.73 % | ||
PTV_6300_EVAL | PTV_6300_EVAL | V59.85Gy[%] | >=99 | 97 | Goal | 99.86 % | ||
PTV_5600_EVAL | PTV_5600_EVAL | V56.0Gy[%] | >=95 | 90 | Goal | 97.44 % | ||
PTV_5600_EVAL | PTV_5600_EVAL | V53.2Gy[%] | >=99 | 97 | Goal | 99.60 % | ||
SPINAL_CORD | SPINAL_CORD | D0.03cc[Gy] | <48 | Goal | 44.531 Gy | |||
SPINAL_CORD_PRV5 | SPINAL_CORD_PRV5 | D1.0cc[Gy] | <=50 | 54 | Goal | 46.501 Gy | ||
BRAINSTEM | BRAINSTEM | D0.03cc[Gy] | <54 | Goal | 52.030 Gy | |||
BRAINSTEM_PRV5 | BRAINSTEM_PRV5 | D1.0cc[Gy] | <=57 | 60 | Goal | 55.319 Gy | ||
PAROTID_L | PAROTID_L | Mean[Gy] | <=26 | 40 | Goal | 25.394 Gy | ||
PAROTID_R | PAROTID_R | V30.0Gy[%] | <=50 | 65 | Goal | 48.86 % | ||
LARYNX | LARYNX | Mean[Gy] | <=40 | 55 | Goal | 25.489 Gy | ||
SUBMND_SALV_L | SUBMND_SALV_L | Mean[Gy] | <=35 | 65 | Variation | 35.401 Gy | ||
COCHLEA_L | COCHLEA_L | D0.03cc[Gy] | <=15 | 40 | Goal | 13.786 Gy | ||
COCHLEA_R | COCHLEA_R | D0.03cc[Gy] | <=15 | 40 | Goal | 13.027 Gy | ||
MANDIBLE | MANDIBLE | V70.0Gy[%] | <=0 | 5 | Variation | 0.41 % | ||
LIPS | LIPS | V30.0Gy[%] | <=0 | 30 | Variation | 0.17 % | ||
POST_NECK | POST_NECK | V35.0Gy[%] | <=10 | 40 | Goal | 6.48 % |
This plan utilized nine (9) fields with manually selected gantry positions to follow the shape of the low dose target in the transverse view. Also, beam edges were placed so a sharp gradient could be achieved against the highest value OARs (parotids, submandibular glad, oral cavity, spinal cord). From a beam's eye view, unique collimator rotations per field were selected to match the general angle of the target volume.
This static field IMRT plan scored better, retrospectivly, than any plan submitted to this 2012 plan challenge.
This 9F IMRT plan for an oropharyngeal primary treated bilaterally shows adequate coverage of the target, with no significant heterogeneities and outside of the target hotspots. Sharp gradients are achieved around all OARs, with all being well within assigned constraints. The sharp gradients are so remarkable that it would be interesting to plan recurrent H&N cases for SBRT with this technique. This is a clinically acceptable plan.
3rd party software plan report |
Dosimetric ScoreCard |
DICOM patient export |
Any reference to a "plan study" are simply what the organizers call each case and may not be a "study" in the FDA sense as they may not have been published in a peer reviewed journal.
Varian does not provide medical advice and these are illustrative examples only.
Leading plans by expert planner. Your results may vary.