Name (ID): P4Brain, P4Brain (P4Brain)
Plan or PlanSum ID: SX2
| Structure ID | Structure Code | Patient Structure | DVH Objective | Evaluator | Variation | Priority | Met | Achieved |
| 4GyBrain_noPTV | BRAIN_PTV | V4Gy[cc] | 76.51 cc | |||||
| 12GyBrain_noPTV | BRAIN_PTV | V12Gy[cc] | 7.89 cc | |||||
| 21GyBrain_noPTV | BRAIN_PTV | V21Gy[cc] | 0.62 cc | |||||
| Brain_noPTV | BRAIN_PTV | Mean[Gy] | 1.422 Gy | |||||
| PTV1 | 1PTV_RF_21 | V100%[%] | >=99 | Goal | 99.09 % | |||
| GTV1 | 1GTV_RF | Max[%] | 133.4 % | |||||
| GTV1 | 1GTV_RF | Max[Gy] | 28.022 Gy | |||||
| GTV1 | 1GTV_RF | Min[Gy] | 23.089 Gy | |||||
| GTV1 | 1GTV_RF | V100%[cc] | 0.31 cc | |||||
| GTV1 | 1GTV_RF | Volume[cc] | 0.31 cc | |||||
| PTV2 | 2PTV_RC_21 | V100%[%] | >=99 | Goal | 99.24 % | |||
| GTV2 | 2GTV_RC | Max[%] | 131.8 % | |||||
| GTV2 | 2GTV_RC | Max[Gy] | 27.683 Gy | |||||
| GTV2 | 2GTV_RC | Min[Gy] | 20.107 Gy | |||||
| GTV2 | 2GTV_RC | V100%[cc] | 0.40 cc | |||||
| GTV2 | 2GTV_RC | Volume[cc] | 0.40 cc | |||||
| PTV3 | 3PTV_LC_21 | V100%[%] | >=99 | Goal | 99.49 % | |||
| GTV3 | 3GTV_LC | Max[%] | 134.3 % | |||||
| GTV3 | 3GTV_LC | Max[Gy] | 28.195 Gy | |||||
| GTV3 | 3GTV_LC | Min[Gy] | 24.332 Gy | |||||
| GTV3 | 3GTV_LC | V100%[cc] | 0.28 cc | |||||
| GTV3 | 3GTV_LC | Volume[cc] | 0.28 cc | |||||
| Brainstem | Brainstem | Max[Gy] | 2.798 Gy | |||||
| Chiasm | Chiasm | Max[Gy] | 1.384 Gy | |||||
| OpticNerve_L | OptNrv_L | Max[Gy] | 1.642 Gy | |||||
| OpticNerve_R | OptNrv_R | Max[Gy] | 1.485 Gy | |||||
| Eye_L | Eye_L | Max[Gy] | 1.482 Gy | |||||
| Eye_R | Eye_R | Max[Gy] | 1.737 Gy |
This plan was created with .5cm effective MLC leaf width utilizing a works-in-progress software configuration. Also of note, this plan doesn't include imaging dose in the final calculation. Otherwise, this plan used simple optimization objectives: max dose constraints to OARs and miniumum dose contraints to targets. The manual NTO was employed with Priority 140, Distance .3cm, Start Dose 100%, End Dose 8%, Fall Off .25. Extended convergence mode was selected in the Photon Optimizater calculation options in Eclipse V15.5+.
|
3rd party software plan report |
DICOM patient export |
Any reference to a "plan study" are simply what the organizers call each case and may not be a "study" in the FDA sense as they may not have been published in a peer reviewed journal.
Varian does not provide medical advice and these are illustrative examples only.
Leading plans by expert planner. Your results may vary.
FOR EDUCATIONAL AND SCIENTIFIC EXCHANGE ONLY – NOT FOR SALES OR PROMOTIONAL USE.