3:44 IMRT treatment field delivery time (estimated)
Name (ID): 2017 QADS Head Neck PS, head and neck (2017 QADS PS HN)
Plan or PlanSum ID: 144.49
Structure ID | Structure Code | Patient Structure | DVH Objective | Evaluator | Variation | Priority | Met | Achieved |
PTV_70 | PTV_70 | V70Gy[%] | >=95 | 90 | Variation | 94.36 % | ||
PTV_70 | PTV_70 | V73.5Gy[%] | <=0 | 10 | Goal | 0.00 % | ||
PTV_70 | PTV_70 | D0.03cc[Gy] | <71.3 | 77 | Variation | 72.670 Gy | ||
CTV_70 | CTV_70 | V70Gy[%] | >=99 | 95 | Goal | 99.79 % | ||
PTV63 - PTV70 | PTV63 - PTV70 | V63Gy[%] | >=95 | 90 | Goal | 95.98 % | ||
PTV63 - PTV70 | PTV63 - PTV70 | V66.15Gy[%] | <=20 | 60 | Variation | 31.19 % | ||
CTV63 - CTV70 | CTV63 - CTV70 | V63Gy[%] | >=99 | 95 | Goal | 99.75 % | ||
PTV56 - PTV63 | PTV56 - PTV63 | V56Gy[%] | >=95 | 90 | Goal | 95.76 % | ||
PTV56 - PTV63 | PTV56 - PTV63 | V58.8Gy[%] | <30 | 60 | Goal | 24.18 % | ||
CTV56 - CTV63 | CTV56 - CTV63 | V56Gy[%] | >=99 | 95 | Goal | 99.56 % | ||
SPINAL_CORD | SPINAL_CORD | D0.03cc[Gy] | <47.999 | 48 | Goal | 47.074 Gy | ||
BRAINSTEM | BRAINSTEM | D0.03cc[Gy] | <51.999 | 52 | Goal | 50.644 Gy | ||
RT COCHLEA | RT COCHLEA | D0.03cc[Gy] | <=35 | 45 | Goal | 24.183 Gy | ||
LT COCHLEA | LT COCHLEA | D0.03cc[Gy] | <=35 | 45 | Goal | 28.486 Gy | ||
LIPS | LIPS | V30Gy[%] | <=20 | 40 | Goal | 15.52 % | ||
RT_PAROTID | RT_PAROTID | Mean[Gy] | <24 | 30 | Goal | 23.782 Gy | ||
MANDIBLE1 | MANDIBLE1 | V70Gy[%] | <=10 | 25 | Goal | 9.16 % | ||
LARYNX | LARYNX | Mean[Gy] | <40 | 50 | Goal | 38.958 Gy | ||
Post Neck1 | Post Neck1 | V35Gy[%] | <=10 | 30 | Goal | 9.14 % |
This plan was created with the inital 1cm effective leaf size configuration. Nine (9) static gantry positions were selected such that the beams follow the inverted V target shape in the transverse view. Uniquie colimator rotations were selected from the beams eye view.
3rd party software plan report |
DICOM patient export |
Any reference to a "plan study" are simply what the organizers call each case and may not be a "study" in the FDA sense as they may not have been published in a peer reviewed journal.
Varian does not provide medical advice and these are illustrative examples only.
Leading plans by expert planner. Your results may vary.
FOR EDUCATIONAL AND SCIENTIFIC EXCHANGE ONLY – NOT FOR SALES OR PROMOTIONAL USE.